GC Image Users' Guide

GCxGC-MS Data

Mass spectrometry (MS) is a useful tool for computer-assisted and automated analyses of the complex separations produced by GCxGC. In GCxGC-MS, each sample output of the GCxGC separation is analyzed by MS, e.g., time-of-flight (TOF) MS. Then, instead of a single value at each sample output, there is a mass spectrum consisting of an array of (mass/z, intensity) pairs. These mass spectra can be used in identifying unknown chemicals and in separating co-eluting peaks for more accurate quantification.

GCxGC-MS File Input and Output

GC Image stores GCxGC-MS data in two files: a GCI file containing metadata (as described in chapter Image Input and Output) and a binary file containing the MS data in the network Common Data Format (NetCDF). NetCDF, developed at the Unidata Program Center and distributed by the University Corporation for Atmospheric Research, is "a machine-independent format for representing scientific data." [www.unidata.ucar.edu] NetCDF format is the basis for the Analytical Data Interchange (ANDI) Protocol for Mass Spectrometric Data [ASTM E2077-00 and E2078-00] and most MS systems support export of data in CDF format for non-proprietary data interchange. Using the GCI and CDF files, the GC Image file operations Open, Recent Images, Save, Save As, Close, and Exit work for GCxGC-MS images as they do for GCxGC images (described in chapter Image Input and Output).

GC Image supports importing GCxGC-MS data from ANDI files. To import GCxGC-MS data contained in an ANDI file (with .cdf extension), select the Import Image option from the File menu or click the Import Image button from the tool bar of the Image Viewer. GC Image prompts the user to enter the source and target filenames and metadata (such as size). GC Image then imports the GCxGC-MS data and displays an image of the total ion count (TIC) at each pixel, as pictured in Figure 1. The TIC is the sum of intensities of all mass/z values for the sample. Most of the image visualization, processing, and analysis tools for GCxGC data are available for GCxGC-MS data. Some of the tools specific to GCxGC-MS are described in this chapter.


Figure 1: An image of the total ion count (TIC) values for GCxGC-MS data.

For GCxGC-MS data, the import dialog allows the user to optionally limit the maximum number of values in each mass spectrum. GCxGC-MS analyses generate very large data sets that may exceed the memory capacity of desktop systems, causing very slow processing. One approach to processing GCxGC-MS data more quickly is to reduce the size of the GCxGC-MS data by retaining only the elements with the largest intensity values in each mass spectrum and using sparse matrices to store the reduced mass spectra.

For example, a GCxGC-MS image with 756x800 GCxGC samples and 215 intensity values in each mass spectrum requires nearly one gigabyte of memory (using four byte integers for the mass/z and intensity values) as shown in Table 1. If only the 32 elements with the largest intensity are stored for each mass spectrum, the memory requirement is reduced drastically to about 0.15 gigabytes as shown in Table 2.

Mass/Z values array756 x 800 x 215 x 4 = 520,128,000 Bytes 496.03 MB
Intensity values array756 x 800 x 215 x 4 = 520,128,000 Bytes 496.03 MB
Total memory required to store both the arrays: 992.06 MB
Table 1. Memory required to store the full mass spectrum for each sample in a GCxGC-MS image.

Mass/Z values array756 x 800 x 32 x 4 =   77,414,400 Bytes   73.82 MB
Intensity values array 756 x 800 x 32 x 4 =   77,414,400 Bytes   73.82 MB
Total memory for both the arrays147.64 MB
Table 2. Memory required to store the 32 values with the largest intensity for each mass spectrum in a GCxGC-MS image.

Of course, the gain in storage and processing efficiency must be weighed against the loss of data. Figure 2 shows the reduction of a mass spectrum to retain only the values with the largest intensity in a sparse representation.

Figure 2: An example of mass spectrum data reduction with sparse representation.

Mass Spectrum Viewer

The GC Image Mass Spectrum Viewer provides both graphical and tabular views of the selected mass spectrum (e.g., as illustrated in Figure 3).

Figure 3: The Mass Spectrum Viewer GUI.

The tabular view (on the right side) has two columns: one for mass/z values and one for the corresponding intensity values in the mass spectrum. The intensity values can be expressed as absolute values, as percentages of total intensity (so that all values sum to 100), or as relative percentage values (so that the largest valued entry is 100.00). The choice of value type is set by the radio buttons at the bottom of the window. The entries in the table can be sorted in forward or reverse order by clicking on the label at the head of the column to be used as the sort key. (Consecutive clicks reverse the sort order.) At the top of the table is the total intensity count (TIC) for the mass spectrum.

In the graph (on the left side of the window), the mass/z values are on the horizontal axis and the intensity values are on the vertical axis. The range for the horizontal axis is determined by the minimum and maximum mass/z values in all mass spectra of the image (so, the same absolute scale is used for the horizontal axis in all the mass spectra graphs). The domain for the vertical axis is from zero and to the largest intensity value in the selected mass spectrum. The vertical axis is expressed as absolute value, percentage of total intensity, or relative percentage, depending on the setting of the radio buttons at the bottom of the window.

As the cursor is moved across the graph, the mouse-over tool-tip indicates the mass/z and intensity value, as in Figure 3. Depressing the control key with click-and-drag of the right mouse button defines a zoom window for redisplay of a region of the graph. To return to a full view of the graph, click the Reset button.

Mass Spectrum Selection Modes

GC Image supports selecting mass spectra for viewing in three modes:

As illustrated in Figure 4, the Image Viewer palette contains a button to set the cursor mode to mass spectrum and a pull-down mass spectrum mode selector.

Figure 4: Mass spectrum mode selection.

In Point Mass Spectrum mode, when the user clicks the left mouse-button on a point in the TIC image (e.g., as in Figure 1), GC Image retrieves the mass spectrum for the corresponding sample and displays the mass spectrum in a popup window, as in Figure 3.

In Peak Mass Spectrum mode, when the user clicks in a blob in the TIC image, GC Image displays the mass spectrum of the sample within the blob that has the largest total ion count (TIC), i.e., the mass spectrum at the blob peak. This mode is accessible only after blob detection has been performed.

In Blob Mass Spectrum mode, when the user clicks in a blob in the TIC image, GC Image displays the sum of all of the mass spectra of all the samples in the selected blob. In computing the sum, the mass/z values are rounded to the nearest integer. This mode is accessible only after blob detection has been performed.

Using the Mass Spec panel in Configure Settings, the user can specify whether successive selections of mass spectra are presented in multiple windows (in which case a new window is opened for each successive mass spectrum) or in a single window (in which case the previous mass spectrum graph is replaced in the same window by the new mass spectrum). The Windows pull-down menu in the Image Viewer allows switching focus between multiple windows that may be open.

Mass Spectrum Ranges

The Mass Spectrum Viewer allows the user to subset the mass spectrum, by selecting one or more mass/z subranges. Mass spectrum subranges are particularly useful for generating ion chromatogram images, described below. To specify a subrange of mass/z values (including a subrange with a single mass/z value), the user clicks-and-drags the left mouse-button in the graph area of the Mass Spectrum Viewer. Additional subranges can be added by depressing the Control-Key during click-and-drag. Overlapping subranges are merged. A click-and-drag without the control key clears any previously selected subranges. Figure 5 illustrates a mass spectrum with two subranges specified.

Figure 5: A mass spectrum in the Mass Spectrum Viewer with two subranges selected.

Ion Chromatogram Images

The Mass Spectrum Viewer has a button to generate an ion chromatogram image. Ion chromatogram images are constructed with each pixel having the value of the total intensities in one or more mass/z subranges in the mass spectrum of the corresponding sample. An example ion chromatogram image for the mass/z subrange 120-134 is illustrated in Figure 6.

To generate an ion chromatogram image, first select a chromatogram (for a pixel, peak, or blob) displayed in a Mass Spectrum Viewer window. Then, in the graphical display area of the Mass Spectrum Viewer, select one or more mass/z subranges. (A subrange may have a single mass/z value.) Finally, click the Open Ion Image button in the Mass Spectrum Viewer control area.

Using the Mass Spec panel in Configure Settings, the user can specify whether successive ion chromatogram images are presented in multiple windows (in which case a new window is opened for each successive ion chromatogram) or in a single window (in which case the previous ion chromatogram is replaced in the same window by the new ion chromatogram image). The Windows pull-down menu in the Image Viewer allows switching focus between multiple windows that may be open.


Figure 6: An example ion chromatogram image for the mass/z subrange 120-134.

Selection Expressions

GC Image provides a tool for constructing and applying selection expressions. A user can build an expression with criteria for selecting pixels or blobs based on retention time and mass spectrum. For example, a simple expression in the selection language is:
Retention(1)>15
This expression could be used to select all pixels with first-dimension retention time greater than 15 minutes or to select all blobs in which the largest-valued pixel (i.e., the peak) has first-dimension retention time greater than 15 minutes.

The tool, pictured in Figure 7, resembles a calculator, with a display box for the expression and buttons for numbers, functions, arithmetic operators, logical operators, relational operators, and parentheses for grouping. The user builds the expression by typing directly in the display box and/or clicking buttons. Buttons for Backspace and Clear also are available to construct the expression.

Figure 7: Selection Expression tool.

Once constructed, the expression is applied either pixel-by-pixel or blob-by-blob, depending on the setting of the radio buttons. In pixel selection mode, the expression is applied by clicking the Generate Selection Image button, which produces an image in which pixels that evaluate to true are unchanged and pixels that evaluate to false are set to zero (i.e., masked). In blob selection mode, the expression is applied by clicking the Select Blobs button, which causes the blobs that evaluate to true to be selected and the blobs that evaluate to false to be un-selected. There are two blob selection modes: peak mode, which uses the mass spectrum of the blob's peak-valued pixel, or integrated mode, which uses the mass spectrum formed by summing the mass spectra of all pixels in the blob.

The functions available for selection expressions are:

The operators available for selection expressions are:

Here is a more interesting example selection expression:

(Retention(2)>2.5)&((Ordinal(128)=1)|(Ordinal(142)=1))
This expression can be used to select all blobs that have a peak with second-column retention time of greater than 2.5 seconds and which have maximum mass spectral intensity at either mass/z=128 or mass/z=142.

Expressions can be stored and recalled with the selection calculator. To save an expression, type a name for the expression in the Name text box before evaluating the expression. To recall an expression, select the expression from the list of named expressions. The user is prompted before overwriting a previously saved expression or reading to replace the current expression.

Mass Spectrum Library Search

The Mass Spectrum Viewer has a button to initiate a library search for the selected mass spectrum. The Lookup Library button invokes a library search to match the selected mass spectrum. A library search also can be performed using mass spectrum subranges.

GC Image interfaces with the National Institute of Standards and Technology (NIST) MS Search Program (Version 2.0) and through it can search the NIST/EPA/NIH Mass Spectral Library, the NIST/EPA/NIH Mass Spectral Demonstration Library, or any mass spectral library that is format-compliant. The CDROM installs the MS Search Program (Version 2.0) and the NIST/EPA/NIH Mass Spectral Demonstration Library, which contains 2400 compounds, a small fraction of the full NIST/EPA/NIH Mass Spectral Library.

GC Image currently supports identity search of the designated library (or libraries). If the unknown compound is likely to be in the library, identity search is the quickest way to find it. Identity search uses a very efficient method of selecting a small set of spectra for subsequent spectrum-by-spectrum comparison. The two types of identity search are Quick Search and Normal Search.

The NIST Mass Spectral Search Program returns a "hit list" of matched chemical compounds from the library and several factors for each matched compound. GC Image displays these in a table, as illustrated in Figure 8. Compounds in the hit list are listed with the following factors:

Figure 8: Example MS Search hit list.

Library Search for Blob Identification

By default, the hit list contains ten compounds with the highest match factor. The user can configure the size of the hit list using the Library Lookup panel in Configure Settings, illustrated in Figure 9. The user also can designate the library to be searched and the type of identity search.

Figure 9: Library search configuration pane.

Library Search for Blob Identification

GC Image can use the library search to automatically set the compound names for blobs. Clicking the Search MS Library button on the Image Viewer tool bar or selecting Search MS Library from the Filter menu submits one or more blob (peak or integrated) mass spectra for library search and sets the compound name of each submitted blob to that of the library compound with the largest match factor for that blob's mass spectrum. After library search for a blob, GC Images maintains the "hit list" for that blob, which can be accessed via the blob metadata popup, as pictured in Figure 10. The user can choose interactively any compound in the hit list as the name of the blob, overriding the initial identification of the compound with the largest match factor, as illustrated in Figure 11.
Figure 10: Blob Metadata popup with a button to access the hit list.
Figure 11: Example MS Search hit list for a blob.

The library search be made on all blobs or on all selected blobs and the search can be made using either the peak mass spectrum or the integrated mass spectrum of each blob (configured by the user, as illustrated in Figure 8). The user also can set the configuration so that library search is performed automatically after blob detection.